A peptide certificate of analysis, usually shortened to COA, is the document that links a physical vial to a set of laboratory measurements.

If you want to know how to read a COA, the short answer is that you are checking four things: that the document describes the same material you hold, that the test methods are the right ones, that the numbers meet a stated specification, and that the laboratory behind it can be identified.

This guide goes through a typical peptide COA line by line. It explains what an HPLC purity figure measures and what it leaves out, how a mass spectrometry result confirms identity, why purity and net peptide content are different numbers, which optional tests you may see, and how to spot a certificate that has been reused or altered.

A worked example uses BPC-157, whose published average molecular weight is 1419.5 g/mol, to show how an observed mass is compared with a theoretical one. A checklist at the end condenses the whole process into a few minutes of checking.

Everything here concerns the analysis of the material as supplied. GBPeptides sells peptides strictly for laboratory and in-vitro research, not for human or animal consumption.

What a certificate of analysis is

A certificate of analysis is a report issued by a testing laboratory, or compiled by a manufacturer from laboratory data, that states which tests were run on one specific batch of a material, which method was used for each, and what was found. It is a record of measurement, not a general statement about a product line.

Three features separate a useful COA from a decorative one. It is batch-specific, so it carries a lot or batch number rather than only a product name. It reports results against a specification, so a reader can see what was required as well as what was found.

And it is traceable, so the laboratory, the date and the method can be identified and, if necessary, queried.

For peptides, the two measurements that matter most are purity and identity. Purity is normally reported from reversed-phase high-performance liquid chromatography (HPLC). Identity is normally reported from mass spectrometry (MS). The two answer different questions, which is why a COA that gives only one of them is incomplete.

The companion guide on HPLC versus mass spectrometry explains the techniques in more depth.

A COA is also a point-in-time document. It describes the batch on the date it was tested. It says nothing about what happened to a vial afterwards, which is one reason storage conditions matter and why the lot number on the vial has to be matched to the certificate.

The header fields: product, lot, dates and laboratory

The top of a certificate identifies what was tested and who tested it. These fields look administrative, but most fake or recycled certificates fail here first.

FieldWhat it should showWhat to check
Product nameThe compound name, and ideally a synonym or codeIt matches the product you hold, including variants such as CJC-1295 with or without DAC
Lot or batch numberA unique identifier for one manufacturing batchIt is identical to the number printed on the vial, character for character
Sample descriptionPhysical form, usually white lyophilised powderIt matches the form you received; GHK-Cu is blue by nature
Date of manufacture or synthesisWhen the batch was madePresent, and earlier than the test date
Date of analysisWhen the tests were runPresent, and consistent with the lot; an undated report cannot be placed in time
Laboratory name and addressThe organisation that ran the testsA real, identifiable organisation, not only a logo
AccreditationA standard such as ISO/IEC 17025, and the accrediting bodyStated accurately; see below
Analyst or approverA named person or signaturePresent; an anonymous report is harder to query
Method referenceInstrument, column, wavelength, mobile phase or a method IDSpecific enough to be repeatable

The product name alone is weak evidence. Plenty of reports are generated from a template with the name typed in. The lot number is the field that ties the document to a particular vial, so regard it as the primary key of the whole certificate.

Dates deserve a moment. A date of analysis that precedes the date of synthesis is an error. A date of analysis several years before the lot was supposedly made is a stronger warning. Equally, a certificate that gives no analysis date at all cannot tell you how fresh the data are.

Laboratory accreditation: ISO/IEC 17025 and UKAS

Many COAs state that the testing laboratory is accredited to ISO/IEC 17025. That standard sets out general requirements for the competence, impartiality and consistent operation of testing and calibration laboratories. Accreditation means an independent accreditation body has assessed the laboratory against the standard.

In the United Kingdom the national accreditation body is UKAS, the United Kingdom Accreditation Service. Laboratories in other countries are accredited by their own national bodies, many of which recognise one another's accreditations through international arrangements. A UKAS-accredited laboratory has a schedule of accreditation published on the UKAS website, listing exactly what it is accredited to test.

That last point is the one most often missed. Accreditation is granted for a defined scope: particular techniques, on particular kinds of material. A laboratory can hold ISO/IEC 17025 accreditation for, say, one chromatographic method on one sample type and still run other tests outside its accredited scope.

A statement that a laboratory is 'ISO accredited' without naming the standard, the body or the scope tells you very little.

Accreditation is not a requirement for a useful COA, and an unaccredited laboratory can produce sound data. But it is a checkable claim. If a certificate names an accreditation, you can look up the accreditation body's directory and confirm that the named laboratory is listed.

If it names an accreditation that the laboratory does not hold, that is a serious credibility problem for the whole document.

Independence is the other part of the picture. A certificate issued by a third-party laboratory that has no stake in the sale carries more weight than one produced by the manufacturer's own bench. GBPeptides has every batch tested by an independent laboratory by HPLC and mass spectrometry before sale.

Specification versus result

A good COA has at least three columns: the test, the specification and the result. The specification is the acceptance criterion set before testing, for example a minimum HPLC purity or an allowed mass window. The result is what was measured.

Reading the two side by side is the most useful habit you can build. A result with no specification is a number without a yardstick. A specification written to match the result after the fact, or one so loose that nearly any material would pass, shows the same weakness from the other side.

For identity by mass spectrometry, a sensible specification is expressed as an expected mass with a tolerance. For purity, it is expressed as a stated minimum.

Also read the units and the basis. 'Purity: 99' means little without 'by HPLC area at 220 nm'. A purity figure can be computed over a defined integration window or over the entire chromatogram, and the two can differ. The method line should tell you which.

Finally, check that the result field contains a result. Some certificates carry only the words 'conforms' or 'pass'. A bare pass is a conclusion without evidence. You want the measured value, and for chromatography you want the chromatogram itself.

HPLC purity: what the number is

HPLC purity is the headline figure on most peptide COAs. The method is reversed-phase HPLC: a solution of the sample is carried by a liquid through a column packed with hydrophobic particles, most often C18-bonded silica.

A gradient that gradually increases the proportion of an organic solvent such as acetonitrile, normally with a small amount of an ion-pairing acid such as trifluoroacetic acid (TFA), causes the components of the sample to leave the column one after another, roughly in order of increasing hydrophobicity.

A detector at the column outlet records absorbance as time passes. For peptides the detector is almost always set at a low ultraviolet wavelength, commonly around 214 to 220 nm, where the peptide bond itself absorbs. This is useful because every peptide has peptide bonds, so every peptide gives a signal regardless of its side chains.

Some reports add a second trace at 254 or 280 nm, where aromatic residues such as tryptophan, tyrosine and phenylalanine absorb.

The output is a chromatogram: a trace with a peak for each component that absorbs at the chosen wavelength. The main peak is the target peptide. Smaller peaks are other things that were in the sample.

The integration software measures the area under each peak, and purity is reported as the main peak's area divided by the sum of all integrated peak areas, expressed as a percentage. This is called area percent, or area normalisation.

On the certificate you should see the chromatogram itself, a peak table with retention time, area and area percent for each peak, and the method conditions. A typical peak table has a dominant main peak and a few small peaks.

The retention time of the main peak is a useful secondary clue: for a given method and column it should be reproducible between batches of the same compound.

  • Retention time: how long after sample loading the peak emerges. It depends on the column, gradient and temperature, so it is only comparable within the same method.
  • Peak area: the integrated signal, proportional to how much absorbing material produced the peak.
  • Area percent: that peak's area as a share of the total integrated area.
  • Wavelength: where the detector was set, typically in the 214–220 nm region for peptides.
  • Gradient and mobile phases: usually water and acetonitrile, each with a small amount of acid such as TFA or formic acid.

When you open the chromatogram, look at it rather than only at the percentage. The main peak should be sharp and reasonably symmetrical. A broad, tailing or split main peak suggests a method problem or a mixture. The baseline should be flat apart from the gradual drift that gradient elution causes at low wavelengths.

Peaks before the main peak are typically more polar species, and peaks after it are more hydrophobic ones; the very first peak is often unretained material from the solvent front and is not part of the sample.

What HPLC purity does not tell you

An HPLC area-percent figure is a relative measure of UV-absorbing components that the method separated. That definition has limits, and a careful reader keeps them in mind.

First, it only counts what absorbs at the chosen wavelength. A component with no chromophore at 214 to 220 nm, or one that absorbs weakly, is under-counted or invisible. Salts, many solvents and some counter-ions do not appear.

Second, it only counts what separated. If an impurity elutes at the same time as the main peak, the two are integrated as one. Impurities that differ from the target by a single amino acid substitution, a small modification or a change of stereochemistry can sit very close to the main peak.

A different gradient, column chemistry or pH can separate them, which is why a single method is not definitive.

Third, area percent assumes every component responds equally at the detector. In practice response factors differ, particularly between a full-length peptide and a short fragment with fewer peptide bonds. The area share is therefore an estimate of the mass share, not an exact figure.

Fourth, it says nothing about identity. A clean single peak is a clean single peak of something. The retention time does not prove which compound it is. That is the job of mass spectrometry.

Fifth, it does not report how much of the vial's contents is peptide. Water, residual salts and counter-ions are not part of an HPLC purity figure. That is covered under net peptide content below.

None of this makes HPLC purity unhelpful. It is the standard, widely accepted way to estimate the relative amount of peptide-related impurities in a batch. It simply answers a narrower question than is sometimes assumed.

Mass spectrometry identity: observed against theoretical

Mass spectrometry confirms identity by measuring molecular mass. The sample is ionised, commonly by electrospray ionisation (ESI) for peptides, or by matrix-assisted laser desorption ionisation (MALDI), and the instrument reports the mass-to-charge ratio, written m/z, of the ions produced. A COA compares the observed m/z with the value expected from the peptide's known composition.

Because the instrument measures mass divided by charge, the observed number depends on how many protons the molecule picked up. A singly charged ion, written [M+H]+, has one added proton, so its m/z is approximately the molecular mass plus 1.007.

A doubly charged ion, [M+2H]2+, has two added protons and carries two charges, so its m/z is approximately (M + 2 × 1.007) ÷ 2. Larger peptides often show several charge states in the same spectrum, for example [M+2H]2+, [M+3H]3+ and [M+4H]4+. Seeing a consistent series, all pointing at the same molecular mass, is stronger evidence than a single peak.

The theoretical mass you compare against comes in two flavours. The average mass uses the weighted-average atomic masses of each element across all isotopes, and corresponds to the molecular weight printed in most catalogues. The monoisotopic mass uses only the most abundant isotope of each element, for example carbon-12 and hydrogen-1.

For small peptides, high-resolution instruments resolve the isotope pattern and report the monoisotopic peak. For larger molecules, lower-resolution instruments often report something closer to the average mass.

That distinction explains why two COAs for the same compound can report slightly different numbers without either being wrong. What matters is that the observed value matches the theoretical value computed on the same basis, within a stated tolerance.

The tolerance depends on the instrument; a high-resolution instrument might agree to a few parts per million, while a simple quadrupole instrument might agree to a few tenths of a mass unit.

Some reports show a deconvoluted spectrum, in which software converts a series of multiply charged peaks into a single calculated neutral mass. If the report gives only the deconvoluted value, check that the raw spectrum is also provided, since deconvolution can be tuned.

What mass spectrometry confirms is that molecules of the expected mass are present. It does not, on its own, prove the correct sequence. Isomers, including peptides with the same residues in a different order or with a stereochemical change, have identical masses.

Tandem MS (MS/MS) fragmentation or amino acid analysis can add sequence-level information, but a routine identity check is a mass match.

Worked example: matching a BPC-157 mass

BPC-157 is a 15-residue peptide with the sequence Gly-Glu-Pro-Pro-Pro-Gly-Lys-Pro-Ala-Asp-Asp-Ala-Gly-Leu-Val. Its molecular formula is C62H98N16O22 and its published average molecular weight is 1419.5 g/mol. The proton mass used below is 1.007.

Using these figures, the expected positions of the two most common ions are calculated as follows.

IonArithmeticExpected m/z
[M+H]+1419.5 + 1.007about 1420.5
[M+2H]2+(1419.5 + 2 × 1.007) ÷ 2 = 1421.514 ÷ 2about 710.8

A COA for a correct BPC-157 batch should therefore show a prominent peak near 1420.5 if it reports the singly charged ion, and a prominent peak near 710.8 if it reports the doubly charged ion. Many spectra show both. If the report gives a deconvoluted neutral mass, it should be close to 1419.5 on the average-mass basis.

On a monoisotopic basis the numbers differ. Computed from the formula C62H98N16O22 with the most abundant isotopes, the monoisotopic neutral mass is about 1418.70, so the singly charged monoisotopic ion appears near 1419.71.

A high-resolution report that lists 1419.71 is consistent with the same molecule as a lower-resolution report that lists 1420.5; the difference is the mass basis, not the material.

Now consider what a mismatch looks like. A dominant peak at 1436.5 would be about 16 mass units above the expected 1420.5, which is the signature of the addition of one oxygen atom, for instance oxidation.

In this particular sequence there is no methionine, cysteine or tryptophan, so the residues most prone to oxidation are absent, which makes a +16 assignment less likely and would call for a closer look. A dominant peak at a completely different mass would indicate a different compound altogether.

The same check works for any catalogue compound. Take the molecular weight from the product page, compute the singly and doubly charged values, and see whether the certificate's peaks land on them. For the TB-500 and GHK-Cu product pages, the molecular weights are listed alongside the sequence.

Remember that GHK-Cu is a copper complex, so the ion observed depends on whether the copper is retained during ionisation and how the report handles the isotope pattern of copper.

Purity versus net peptide content

Two different numbers are often confused. HPLC purity describes how much of the UV-absorbing peptide material is the target peptide as opposed to peptide-related impurities. Net peptide content describes how much of the vial's total contents, by mass, is peptide at all.

A lyophilised peptide is not a pure dry powder of peptide alone. It also contains residual water, because lyophilised peptides are often hygroscopic and take up moisture from air. It contains counter-ions: a peptide with basic residues such as lysine or arginine, and a free N-terminus, carries positive charges that are balanced by anions.

After purification by HPLC with TFA in the mobile phase, the counter-ion is commonly trifluoroacetate. Peptides supplied as acetate or hydrochloride salts carry those anions instead. These counter-ions are real mass in the vial but invisible to a 214–220 nm UV purity figure.

So a batch can have a very high HPLC purity and a net peptide content that is noticeably lower, because part of the powder mass is water and counter-ion. The gap between the two numbers is usually larger for peptides with many basic residues, since each positive charge needs a counter-ion.

Net peptide content is measured by methods such as amino acid analysis, nitrogen determination or quantitative NMR. It is not a routine entry on every COA, and a certificate that lists only HPLC purity is not hiding anything by omitting it, provided it does not imply that the figure describes the whole vial.

The thing to watch for is a COA, label or listing that presents HPLC purity as though it were the proportion of the vial that is peptide. They are different quantities with different methods, and this guide gives no figure for how large the difference is because it varies by sequence and by batch.

For laboratory work where an exact amount of peptide matters, net peptide content is the relevant number, and the counter-ion form should be known. If the certificate does not state it, the correct inference is that it has not been measured.

Optional tests you may see on a COA

Beyond HPLC and MS, some certificates include additional tests. None is mandatory, but each answers a specific question, and it helps to know what that question is.

TestWhat it measuresWhy it appears
Water content by Karl Fischer titrationResidual water in the powderWater contributes to vial mass and can affect stability; it is one part of the gap between purity and net peptide content
Amino acid analysis (AAA)The ratio and amount of each constituent amino acid after hydrolysisConfirms composition and, with a standard, can give net peptide content
Endotoxin by LAL assayBacterial endotoxin, using a limulus amebocyte lysate reactionRelevant to cell-based work, where endotoxin can interfere with assays
Residual solventsTraces of solvents left from synthesis and purificationTypically by gas chromatography; reports the specific solvents sought
Counter-ion contentTFA, acetate or chlorideShows which salt form the peptide is supplied as
AppearanceVisual description of the powderA simple but useful consistency check; it should match the vial
SolubilityWhether the material dissolves clearly in a stated solventA basic quality check on the physical state of the batch
Peptide sequence by MS/MSFragment masses that map the order of residuesAdds sequence-level evidence beyond a whole-molecule mass

Read extra tests as information, not as decoration. A line reading 'endotoxin: pass' with no method, limit or value is as thin as 'purity: pass'. Equally, the absence of an optional test is not a defect. It is a scope decision, and the document should be judged on whether the tests it does contain are specific and credible.

Matching the COA to the vial in your hands

A genuine certificate for the wrong lot is no use to you. The check takes under a minute. Read the lot number printed on the vial label, find the certificate for that lot, and compare the two strings character by character.

Confirm the product name, and for products that come in more than one size, that the certificate applies to the size you hold.

GBPeptides publishes the certificate for each batch on the product page, a size cannot be bought until its COA is published, and the lot number on the vial matches the COA. When a new lot is supplied, the lot number on the vial changes and so does the certificate it should be compared with.

Keep a record of the lot and the certificate version alongside the date you received the vial.

If the number on the vial does not appear on any published certificate, regard that as unresolved rather than as a clerical detail. Ask the supplier for the certificate that matches the lot, and do not rely on a certificate for a different lot of the same compound as a substitute.

Signs of a reused or fake COA

Certificates can be wrong in several ways, from honest template errors to outright fabrication. The following signs, alone or together, justify asking questions.

  • The same chromatogram appears on certificates for different compounds or different lots. Real batches show small differences in baseline noise, minor peaks and retention time; identical traces are a red flag.
  • The retention time or mass on the certificate does not match the compound. A mass spectrum whose main peak sits nowhere near the expected m/z cannot be explained by the label.
  • The lot number is missing, generic, or does not follow any consistent pattern between certificates from the same source.
  • The laboratory is unnamed, has no address, or the name cannot be found anywhere independent of the document.
  • The accreditation claim cannot be verified in the accreditation body's directory, or names a standard that does not apply to the testing described.
  • The report is a cropped image or screenshot with no way to see the full page, header or method section.
  • Results are stated as a bare 'pass' or 'conforms' with no measured values and no spectrum or chromatogram.
  • Dates are inconsistent: analysis before synthesis, or a date far from the lot's supposed manufacture.
  • The purity value is a suspiciously round number on every certificate from a supplier, or is identical across compounds that were synthesised by different routes.
  • Fonts, alignment or logos differ within one page, suggesting that values were edited into an existing document.
  • The certificate is generic for a product, with no lot, and is reused year after year.

No single item proves fraud. Instruments do produce similar chromatograms for similar compounds, and a template is a normal way to produce reports. But a certificate that fails several of these checks at once is not good evidence of what is in the vial.

The right response is to ask the supplier to provide the original laboratory report, and to decide on that basis.

A practical COA checklist

Run through the following when a certificate arrives. If you cannot tick an item, note it and ask.

  • The lot number on the vial matches the lot number on the certificate exactly.
  • The product name and form match what you received.
  • The laboratory is named, with an address, and any accreditation claim names the standard and body.
  • The date of analysis is present and sensible.
  • The HPLC method is described: column type, wavelength, mobile phase and gradient.
  • The chromatogram is attached, with a sharp main peak and a peak table showing retention time, area and area percent.
  • The purity figure is stated with its basis, for example area percent at a named wavelength.
  • The mass spectrum is attached, and the observed m/z matches the expected value for the stated ion, [M+H]+ or [M+2H]2+, on a consistent mass basis.
  • Results appear as measured values next to a specification, not only as pass or fail.
  • You know what the certificate does not cover, such as net peptide content or counter-ion form.
  • The document is complete, legible and consistent in layout.

A certificate that clears these checks has told you that the batch was measured by recognised methods and that the measurements fit the compound. It has not told you everything, and no COA can. Used well, it is the most reliable piece of evidence that comes with a vial.

Storing the material after you have checked the COA

A COA describes the batch on the day it was tested. Keeping it in that condition depends on storage. Lyophilised peptides as supplied are stored at −20 °C, dry and away from light. The full reasoning, including why a vial should come to room temperature before it is opened, is covered in the guide to storing lyophilised peptides.

If a vial's appearance has changed since you received it, for example the powder has collapsed, discoloured or become sticky, compare it with the sample description on the certificate and record what you see. The certificate cannot tell you about changes that happened after analysis.

Frequently asked questions

What is a peptide certificate of analysis?

A peptide certificate of analysis is a laboratory report for one specific batch. It records which tests were run, which methods were used and what was measured, normally HPLC for purity and mass spectrometry for identity. It should carry a lot number, a test date and the name of the laboratory, so it can be matched to a particular vial.

What HPLC purity should a peptide COA show?

A COA should show a purity value alongside the specification it was tested against, the chromatogram and the method, including the detection wavelength. This guide does not set a threshold, because what counts as acceptable depends on the intended laboratory use. The figure itself matters less than whether it is clearly defined and backed by a chromatogram.

Is HPLC purity the same as net peptide content?

No. HPLC purity is the share of UV-absorbing peptide material that is the target peptide. Net peptide content is the proportion of the vial's total mass that is peptide, once water and counter-ions such as TFA or acetate are accounted for. A batch can have a high HPLC purity and a lower net peptide content.

How do I know a COA matches my vial?

Compare the lot number printed on the vial label with the lot number on the certificate, character by character. Also confirm the product name and, where relevant, the size. At GBPeptides the lot number on the vial matches the COA published on the product page. If the numbers differ, ask for the certificate that matches your lot.

What does [M+H]+ mean on a mass spectrum?

[M+H]+ is the singly charged ion formed when a molecule of mass M picks up one proton, so its m/z is roughly M plus 1.007. For BPC-157, with an average molecular weight of 1419.5, it appears near 1420.5. A doubly charged ion, [M+2H]2+, appears at about half the mass plus one proton, near 710.8.

Does ISO/IEC 17025 accreditation guarantee a good COA?

Not by itself. ISO/IEC 17025 accreditation shows that an independent body has assessed a laboratory's competence, but only for a defined scope of tests. In the UK that body is UKAS. A certificate can still be incomplete or wrong, so check the scope, the method and the data as well as the accreditation claim.

How can I tell if a COA is fake or reused?

Warning signs include an identical chromatogram across different compounds or lots, a mass that does not match the compound, a missing or generic lot number, an unnamed laboratory, a bare pass with no data, and inconsistent dates. One sign proves nothing, but several together justify asking the supplier for the original laboratory report.

Why do some COAs list mass spectrometry values that differ slightly from the molecular weight?

Because of the mass basis and the charge state. Catalogue molecular weights are average masses, while high-resolution instruments often report monoisotopic masses, which are slightly lower. The ion observed also includes added protons. What matters is that the observed value matches the theoretical one calculated on the same basis, within the stated tolerance.